ME12P

Multi-Channel
Electrocardiograph

Is a new type of 12-channel ECG which has made a lot of improvement in the key ECG technologies of quickness, accuracy and stability as well as the streamlined workflow. It applies a new platform for ECG signal acquisition, which can perform well in interface reduction, fast waveform display and right level of filtering.

Features

» 12-channels ECG
» 8.9 inch TFT LCD touch screen
» 60 degree viewing angle adjustable
» Full alphanumeric waterproof silicone keyboard

» Abundant Peripheral interfaces: VGA, LAN, USB, SD CARD
» ECG data management software
» High resolution Printer A4 paper size
» Storage: 1500 ECG Files
» Rechargeable Li-ion battery, 14.8V-4400mAh
Technical Specifications
ECG interpretation
ME12P Digital Electrocardiograph ECG analysis program for adults and pediatrics
Acquisition mode
Simultaneous standard 12-lead in pre-acquisition or post-acquisition
A/D conversion
24 bit
Sampling rate
8,000 samples/s
Frequency response
0.05 Hz~250 Hz
Common mode rejection ratio
>105dB
Polarization voltage
± 550 mV
Input circuit
Floating circuit input
Input impedance
≥50 MΩ
Input CIR current
≤0.1 μA
Patient leakage current
<10 μA
Time constant
≥3.2 s
Noise level
≤15 μVp-p
Sensitivity threshold
≤20 μV
Calibration voltage
1 mV±5 %
Filter setting
EMG filter: 25 / 35 / 75 / 100 / 150 / 250Hz Baseline wander filter: 0.01 / 0.02 / 0.05 / 0.35 / 0.5 / 0.8Hz
AC filter
50Hz, 60Hz
Heart rate range
30~300 bpm
Acquisition time
10~24 seconds
ECG interpretation
ME12P Digital Electrocardiograph ECG analysis program for adults and pediatrics
Acquisition mode
Simultaneous standard 12-lead in pre-acquisition or post-acquisition
A/D conversion
24 bit
Sampling rate
8,000 samples/s
Frequency response
0.05 Hz~250 Hz
Common mode rejection ratio
>105dB
Polarization voltage
± 550 mV
Input circuit
Floating circuit input
Input impedance
≥50 MΩ
Input CIR current
≤0.1 μA
Patient leakage current
<10 μA
Time constant
≥3.2 s
Noise level
≤15 μVp-p
Sensitivity threshold
≤20 μV
Calibration voltage
1 mV±5 %
Filter setting
EMG filter: 25 / 35 / 75 / 100 / 150 / 250Hz Baseline wander filter: 0.01 / 0.02 / 0.05 / 0.35 / 0.5 / 0.8Hz
AC filter
50Hz, 60Hz
Heart rate range
30~300 bpm
Acquisition time
10~24 seconds
Display type
8.9 inch TFT LCD with touch screen
Display resolution
1280×768
AC power supply
AC / DC 100 V~240 V, 50 Hz /60 Hz, 110 VA
Battery power supply
Rechargeable lithium battery, 14.8V, 4400mAh
Recorder resolution
8 points/mm (perpendicular), 40 points/mm (horizontal, 25 mm/s)
Recorder speed
(5, 6.25, 10, 12.5, 25, 50) mm/s±5 %
Recording paper
Z- fold paper: 210mm×150mm×140p
Measurement values
HR, PR interval, QRS duration, QT/QTC interval, P/QRS/T axis, RV5/SV1 voltage
DEMO mode
Normal ECG, Arrhythmia ECG
QTC formula
Bazett, Fridercia, Framingham, Hodges
Data format
ECG, DICOM, XML, JPEG, PDF
External peripheral use
Patient cable socket
SD card port
Two USB port
LAN port
VGA port
Dimensions and weight
Size
345mm x 260mm x 128.5 mm
Weight
About 3.4 kg
Pacemaker Detection
Through ECG waveforms display and analysis, doctors can determine the pacemaker type and work status through the ECG analysis.
Preview Function
Doctors can preview the waveforms and analysis results directly on the screen then decide whether to print or not, good for faster treatment and cost-saving.
R-R Analysis
*30-300 seconds for R-R analysis, better in arrhythmia diagnosis
*Up to 200 kinds of arrhythmia analysis
Waveform Freezing
*300 seconds for waveform freezing and cine-loop, any 10 seconds selectable for analysis and print.
*lt is significant when testing Children, fracture patients, aged people whose skin is too dry to connect the electrodes and so on.